Верификации дозовых распределений, полученных на системе планирования «Гаммаплан»
Т. Г. Ратнер, И. М. Лебеденко, Ю В Журов et al. · CyberLeninK (CyberLeninka) · 2008
The comparison between calculated and measured dose distributions for open and blocked fields was made. It was done for 60Co dose delivery system ROCUS-M within the framework of quality assurance program which is a part of RTPS acceptance tests. It was shown that the data are in a good agreement for all examined fields except for those with «off-axis block». In this case measured profile data slightly exceed the calculated data within the block penumbra region at the margin of open and blocked parts of the field. It was concluded that RTPS GammaPlan with updated calculation technique can be introduced into clinical practice for planning of treatment by 60Co beams both for open and blocked fields.