Objective: To evaluate the biomechanical stability of three-point pedicle screw technique for atlantoaxial instrumentation. Method: Six fresh human cadaveric cervical spines (C0-C5) (age. 25-36 years) were used. After testing the intact spines (group A) instability was induced by removing the anterior arch of atlas, transverse and ajar ligaments and posterior capsule of atlantoxial joints. Then the instable spines were subjected to the following five different techniques, Brook wire fixation technique (group B) C1-C2 left pedicle screws fixation (group C), C1 bilateral pedicle screws and C2 left pedicle screw fixation (group D), C1 left pedicle screw and C2 bilateral pedicle screws fixation (group E). C1-C2 bilateral pedicle screws fixation (group F). Range of motion at C1-C2 was measured at each specimen on six loading directions through spinal flexible motion testing system. The data were analyzed using SPSS software. Result: All 5 techniques significantly decreased range of motion of C1-C2 compared with the intact spine (P<0.05). There was no statistical difference between group D and group E (P>0.05). There was no statistical difference among group D. E. F (P>0.05) in flexion/extension. axial rotation and right lateral bending, however they showed superior to group B and C (P<0.05). As for left lateral bending, group D or E demonstrated a trend toward increasing range of motion compared with group F (P<0.05) but still superior to group B and C (P<0.05) Conclusion: Fixation of atlantoaxial complex using three-point pedicle screw technique can provide reliable stability, which can be used in clinical practice.