在线浓缩柱富集-光学检测-流动注射联用测定海水中痕量铅

邹玉权, 张新申, Hua Li et al. · 2006

A simple, accurate, sensitive and selective flow injection (FI) system for the determination of Pb(Ⅱ) at trace level with a preconcentration step and spectrophotometric detection was proposed. It is based on the preconcentration of lead ions on precolum and dibromo-methyl sulfonazo-Jead complex formation in 1.2 mol/L nitric acid medium with sensitive absorption peak at 630 nm. The experimetnal conditions were optimized. The method provided a liner range between 0.5 to 500 μg /L, a detection limit of 0.1 μg/L and a sample throughout of 24 h. The relative standard deviation was 2.5% for 1.0 μg /L. 1.1% for 5.0 μg /L, lead (Ⅱ) (n=11). The method has been applied to the determination of lead in natural, tap water and seawater samples. and the results of spiked samples are in good agreement with those of inductively coupled plasma-atomic emission spectrometry.

Cited by 0 works.