Разработка и исследования экспериментального образца измерителя концентрации метана для угольных шахт

Вовна Александр Владимирович, А.А. Зори · Известия Южного федерального университета. Технические науки · 2014

The experimental sample of methane concentration meter for coal mines was developed and studied. During laboratory tests in the laboratory aerogas protection were determined sensitivity, noise level and value of basic absolute error of methane concentration measurement. Sensitivity to methane measuring channel meter is not less than 0.17 V/ vol.% in the changing range from 0 to about 2.5 vol.%. Amplitude value of the noise component of the output voltage does not exceed ±12.5 mV, which corresponds to a confidence level of Р=0,95 of the mean square value of ±6.3 mV. It has been established that value of the methane concentration measurement basic absolute error in the specified variation range is not more than ±0.04 vol.% with the required measurement absolute error which specified in the technical specifications for the meter development is no more than ±0.2 vol.%. The dust compensation channel sensitivity of methane meter is 0.024 V / vol.% in the changing range from 0 to 2.5 vol.% was defined. The basic absolute error value of methane concentration measurement with the dust compensation and the dust optics measuring channels value is not more than about ± 0,043 vol.% at required value which is not more than ±0,2 vol.%.