다양한 설계변수를 고려한 치과용 임플란트의 즉시하중과 지연하중의 비교연구
주민진, 전흥재, 한종현 · 대한기계학회 춘추학술대회 · 2009
In this study, the comparison among the calculated stresses on fixtures and their surrounding bones for different implant lengths, diameters and type prior to and after the phase of osseointegration was conducted. A three dimensional finite element analysis was used to determine stresses generated in the bones with the different implant design factors. Twelve FE models, including two designs (tapered, straight) and various dimensions (lengths: 8.5㎜, 10㎜, 11.5㎜, diameters: 4.3㎜, 5.3㎜), with perfectly bonding (delayed loading) and frictional contact (immediate loading) of implant-bone interfaces, were analyzed. The results of this study showed that increasing implant length and diameter result in stress reductions on the bone during delayed loading as well as immediate loading. However, after the length of implant has increased more than a critical length, the excessive stress may generate to the bone for immediate loading. The straight implant showed higher stress levels than tapered implant because of larger surface.