나노유체의 열전도율 측정을 위한 새로운 비정상열선법 센서설계와 자료처리방법
이신표, 이명훈, 김민태 et al. · 대한기계학회 춘추학술대회 · 2003
A fine hot-wire is used both as a heating element and a temperature sensor in transient hot-wire method. The traditional sensor system is unnecessarily big so that it takes large fluid volume to measure the thermal conductivity. To dramatically reduce this fluid volume, a new sensor fabrication and a data processing method are proposed in this article. Contrast to the conventional and most popular two wire sensor, the new sensor system is made up of divided multiple long and short wires. Through validation experiments, it is found that the measured thermal conductivities of the glycerin are exactly same each other between the conventional and proposed new method. Also some technical considerations in arranging the multiple wires are briefly discussed.
Cited by 0 works.