유한요소법을 이용한 치아 임플랜트 피로수명 예측
한인숙, 손정훈, 양영수 et al. · The Journal of Korean Academy of Prosthodontics · 2005
A dental implant plays a role of real teeth but it suffer from both static load and dynamic load. It causes loosening of abutment screw and eventually it could lead a dental implant to fracture because of fatigue. In this paper, using three-dimensional finite element method we studied stress analysis and fatigue life prediction of the Branemark Type dental implant of using the dental service. Finite element analysis and fatigue test were concerned with ISO rules and tensile stress of screw was considered. Using this result, we estimated the fatigue life of the dental implant based on stress-life curve of Titanium. In order to ensure reliability of calculated fatigue life, fatigue test is performed. The calculated and experimental results showed good agreement