산소감응입자를 이용한 마이크로채널 내부의 용존산소 농도장/속도장 동시 측정
김현동, 이승재, 정원택 et al. · 대한기계학회 춘추학술대회 · 2011
An OSParticles-PIV hybrid system is the technique which can be used to measure the oxygen concentration and velocity fields simultaneously using oxygen sensitive functional particles as PIV tracers. In this study, OSParticles were fabricated by dispersion polymerization method, which is one of the chemical synthesis processes, for using the particles as oxygen sensors and PIV tracers. The oxygen concentration and flow velocity simultaneous measurement were carried out in the Y-shaped microchannel by injecting water samples with two different DO concentrations using a double loading syringe pump. Water samples having DO values of 0% and 100% were prepared prior to the experiments. Mercury lamp with 377 nm band-pass filter was used as a light source and images were stored using a CMOS high speed camera (PCO 1200hs) with an Olympus 10x objective lens. The luminescence was filtered using a 550 nm long-pass filter. The velocity field was obtained by the time resolved micro-PIV technique.
Cited by 0 works.