부식피트 및 코팅층을 갖는 치과용 임플란트의 피로수명의 유한요소해석

김민건, Ji Hun Lee, 김요셉 · 대한기계학회 춘추학술대회 · 2009

The body environment causes the generation of corrosion pit on the surface of dental implant and it brings a stress concentration. Therefore, the body environment reduces the fatigue life of implant. Also the hardened layer coated at the surface of abutment screw brings a large stress concentration. The reaction of a body environment like a electrical effect, bit force and body fluid makes a corrosion by breaking away the resistance oxide film from the fixture. Finally it leads dental implant's fracture by a fatigue. In this paper, the evaluation of the fatigue life using the three-dimensional finite element method for stress analysis has been made by FEMFAT. At that, it was concerned with ISO 14801 rules. The estimating the fatigue life is based on stress-life curve of pure Titanium.

Cited by 0 works.

Authors