감성마우스에서 측정된 생체신호의 신뢰성 평가

김현, 허창욱, 최진형 · 한국정신과학학회지 · 2001

We evaluate the reliability of the emotional function BIOPIA Co., Ltd) through comparing the general utilized system MP100 (Biopac Systems Inc.). The 40 Hz re-sampled Photo-Plethysmo Graph (PPG) signals that acquired with the emotional function mouse are processed to determine heartbeat rate with peak detection algonthms (FIR band-pass filter, 6-point difference filter and threshold assign) and tested to how that rate are rehable compare to the MPlOO's one In addition, the correlation of the I Hz re -sampled Galvanic Skin Response(GSR) signals and the MPlOOs GSR signals are tested The evaluation tests are not carried out only In the stable humans condition but In stimulated conditions that caused Continuous Performance Test(CPT) and skipping. In this work, we evaluate the signals that acquired with emotional function mouse and the MP100 systems are strongly correlated over 0.97 in the case of PPG and over 0.98 in case of GSR Moreover, It should be available that Using the emotional function mouse as a utilized bio-Signal acquisition device. mouse(INNO Mouse, bio-signal acquisition