시멘팅 시간에 따른 시멘트와 치환물 사이의 결합강도

조우신, 김병관, 우제호 et al. · 2008

Purpose: The objective of this study was to compare the bonding strength between cement and an implant according to the cementing time. Materials and Methods: The two types of cement used were CMW ® and 3 (Depuy Ltd., Blackpool, UK). Plastic molds containing CMW ® 1 were pressed onto metal blocks at 2, 4, 5 or 6 minutes after mixing the cement, while molds containing CMW ® 3 were pressed onto blocks at 3, 5, 6 or 7 minutes after mixing the cement. Tensile strength was tested with using an Instron Model 8874 (Instron Corp., Canton, MA, USA). Tensile strengths were compared using the Kruskal-Wallis test and the Mann-Whitney U test. Results: The strongest bonding strengths for the tensile load were at 2 minutes post-mixing for CMW ® 1, and at 3 minutes post-mixing for CMW ® 3. The strength rapidly decreased after 5 minutes for CMW ® 1 and after 6 minutes for CMW ® 3. Conclusion: This study suggests that the risk of loosening between cement and an implant is likely to be minimized by the surgical technique that considers the bonding strength according to time.