임플란트의 체결방법에 따른 풀림현상의 역학적 해석에 관한 연구
이범현, 전흥재, 한종현 et al. · 대한기계학회 춘추학술대회 · 2005
A comparative study on the effect of screw loosening of two types of implant systems with different connections was conducted. Three dimensional non-linear finite element analysis is adopted to study the characteristics of screw loosening between external and Internal types of Implant system under the same loading and boundary conditions. The results show that the internal types of implant systems have higher resistance to screw loosening than that of the external types of implant systems since the internal types of implant systems generated larger preload than that generated by the external types for the same tightening moments.
Cited by 0 works.