ВЛИЯНИЕ ДИНАМИЧЕСКОЙ ФИКСАЦИИ ПОЯСНИЧНЫХ ПОЗВОНОЧНЫХ СЕГМЕНТОВ НА ИХ ПОДВИЖНОСТЬ

Александр Евгеньевич Симонович, Маркин Сергей Петрович, Хусниддин Адашалиевич Нуралиев et al. · Russian Journal of Spine Surgery (Khirurgiya Pozvonochnika) · 2008

Objective. Comparative assessment of the influence of dynamic implants COFLEX, DIAM, and DYNESYS on a mobility of lumbar spinal segments in vitro. Material and Methods. Fifteen unfreezed human spine specimens each including L2—L3, L3—L4, and L4—L5 segments were studied. Muscle tissues were removed, while ligaments, facet joints, and intervertebral disc were retained. Segmental motion was tested using a device equipped with tensometric sensors recording the applied force and linear motion of vertebra along three mutually perpendicular axes. The range of linear flexion/extension and lateral motions of vertebrae was measured under the load of -120 to +120 N, and the size of neutral region was determined. Measurements were performed at intact segment after resection of posterior supporting complex, and after placement of dynamical implants. Results. The range of linear flexion/extension motion was reduced by 57 % in specimens with COFLEX, by 32 % — with DIAM, and by 69 % — with DYNESIS implants. All three devices decreased the volume of flexion in a sagittal plane — by 70, 57, and 87 %, respectively. Linear extension was significantly reduced by COFLEX and DYNESIS, and insignificantly — by DIAM. Essential decrease in motion range in lateral bending (25 %) was detected only in specimens with DYNESIS implant. Conclusion. Interspinous implants COFLEX and DIAM limit the motion of spinal segment only in a sagittal plane, though transpedicular system DYNESIS reduces it in sagittal and coronal planes.

Cited by 0 works.

Authors