인공관절 라이너의 내마모성 향상을 위한 초고분자량 폴리에틸렌의 감마선 조사에 따른 경도변화
김현묵, 김동훈, 황창선 et al. · 대한기계학회 춘추학술대회 · 2008
Rockwell hardness test was conducted on conventional and crosslinked ultra-high molecular polyethylene(UHMWPE) all prepared from the same lot of medical grade GUR 1050. The conventional materials were unirradiated and gamma irradiated with 25kGy~200kGy. Gamma irradiated processing was found to significantly impact the crystallinity, and hence the mechanical behavior, of the highly crosslinked UHMWPE. The crystallinity and radiation dose were key predictors of the hardness, plastic flow, and failure properties of conventional and highly crosslinked UHMWPE. The correlation model from experiments would be the basic information to design the liner of artificial joint.
Cited by 0 works.