Reference materials and commutability.
Hubert W. Vesper, W. G. Miller, Gary L. Myers · PubMed · 2007
Maintaining accurate laboratory measurements over time is crucial for assuring appropriate patient care and disease management. Accurate results over time and location are achieved by standardising measurements and establishing traceability to a reference system. Reference materials are key components of such reference systems and for establishing traceability. Commutability of reference materials is a critical property to ensure they are fit for use. Commutability is defined as the equivalence of the mathematical relationships between the results of different measurement procedures for a reference material and for representative samples from healthy and diseased individuals. This material characteristic is of special importance for measurement procedures that are optimised for measuring analytes directly in patient samples. The commutability of a reference material is measurement procedure specific and its assessment requires special experimental designs. This review explains the importance of commutability and summarises different experimental approaches described in the literature that have been used to assess the commutability of reference materials in clinical chemistry.
Cited by 127 works.
Authors
References
- Summary of the second report of the National Cholesterol Education Program (NCEP) Expert Panel on Detection, Evaluation, and Treatment of High Blood Cholesterol in Adults (Adult Treatment Panel II)
- A New Biometrical Procedure for Testing the Equality of Measurements from Two Different Analytical Methods. Application of linear regression procedures for method comparison studies in Clinical Chemistry, Part I
- The accuracy of laboratory measurements in clinical chemistry: a study of 11 routine chemistry analytes in the College of American Pathologists Chemistry Survey with fresh frozen serum, definitive methods, and reference methods.
- A new challenge for America: the National Cholesterol Education Program.
- Evaluation of regression procedures for methods comparison studies
- Glycated Hemoglobin Standardization – National Glycohemoglobin Standardization Program (NGSP) Perspective
- Why Commutability Matters
- Protein Standardization IV: Value Transfer. Procedure for the Assignment of Serum Protein Values from a Reference Preparation to a Target Material
- Accuracy verification and identification of matrix effects. The College of American Pathologists' Protocol.
- Selection, Preparation, and Characterization of Commutable Frozen Human Serum Pools as Potential Secondary Reference Materials for Lipid and Apolipoprotein Measurements: Study within the Framework of the Dutch Project “Calibration 2000”