정적압축하중 하에서 임플란트 시스템의 파손 및 파괴 거동

김경난, 최광훈, 오환섭 et al. · 대한기계학회 춘추학술대회 · 2010

Dental implant system have gotten treatment again, because failure of dental implant system frequently occurred during dental implant treatment or use. Therefore, the aim of this study was to evaluate causes of failure. So we investigated deformation behavior by using load-displacement curve of fixture-abutment combination, and failure mechanism of implant system about tightening torque. As a result, (1)deformation behavior under bending moment didn't combine a system of solid type in lower tightening torque. (2)Fracture of the tightening bolt started the root of the first thread by tensile plastic crack, and then compressive plastic crack occurred at the opposite side. After these two cracks grew, they should finally meet near the center. (3)Plastic crack should grow by Von-Mises stress in stress concentration zone more than by maximum stress in the root of thread.

Cited by 0 works.

Authors