하반신 마비환자를 위한 동력보행보조기의 퍼지제어 기법 개발
강성재, 류제청, 김규석 et al. · Journal of Institute of Control Robotics and Systems · 2009
In this study, we wolud be developed the fuzzy controlled PGO that controlled the flexion and the extension of each PGO’S hip joint using the bio-signal and FSR sensor. The PGO driving system is to couple the right and left sides of the orthosis by specially designed hip joints and pelvic section. This driving system consists of the orthosis, sensor, control system. An air supply system of muscle is composed of an air compressor, 2-way solenoid valve (MAC, USA), accumulator, pressure sensor. Role of this system provide air muscle with the compressed air at hip joint constantly. According to output signal of EMG sensor and foot sensor, air muscles and assists the flexion of hip joint during PGO gait. As a results, the maximum hip flexion angles of RGO’s gait and PGO’s gait were about 16° and 57° respectively. The maximum angle of flexion/extention in hip joint of the patients during RGO’s gait are smaller than normal gait, because of the step length of them shoes a little bit. But maximum angle of flexion/extention in hip joint of the patients during PGO’s gait are larger than normal gait.
Cited by 0 works.