점도 측정표준 불확도 평가
최해만, 윤병로, 이용봉 et al. · 유체기계 연구개발 발표회 논문집 · 2005
Viscosity measurement standards were evaluated according to ISO/IEC 17025. The step-up procedure was employed to calibrate a series of capillary type master viscometers. Uncertainty was calculated with evaluation of various uncertainty factors affected in viscosity measurement. The maximum expanded uncertainty(U) of the master viscometer was 3.0×10?³(at the confidence level of 95 %). This evaluation example will be useful in viscosity measurement uncertainty determination of other standard measurement.
Cited by 0 works.