Исследование влияния биомеханических свойств роговицы на показатели тонометрии

Аветисов Сергей Эдуардович, Бубнова Ирина Алексеевна, Антонов Алексей Анатольевич · Сибирский научный медицинский журнал · 2009

The accuracy of intraocular pressure (IOP) measurement depends on the central corneal thickness, corneal curvature and its viscoelastic properties. The aim: to establish correlation between intraocular pressure indices and biomechanical properties of the cornea and clinical value of the latter for diagnostics of glaucoma. The IOP was measured in 268 eyes of 150 patients with primary openangle glaucoma (POAG), normaltension glaucoma (NTG) and glaucoma suspects by the bidirectional applanation of the cornea, the Maklakov's tonometer, noncontact tonometer (NCT) and electronic tonography. With increasing of the central corneal thickness Goldmann-correlated IOP increased on 4.5 mm Hg on average, IOP measured with NCT increased on 6.2 mm Hg, corneal-compensated IOP (IOPcc) was stable and IOP measured with Maklakov's tonometer increased on 3.1 mm Hg. The difference between IOP measured with the Maklakov's tonometer and IOPcc in patients with «plate» corneas was reliable. We established negative correlation between results of the electronic tonography and the corneal curvature. In our opinion, keratopachymetry, ophthalmometry and bidirectional applanation of the cornea with ORA should be performed in patients with the POAG, NTG and glaucoma suspects.