합성 숙련도 시험시료를 이용한 공기희석 관능법에 의한 복합악취의 숙련도시험평가

김선태, 이석준, 홍석영 et al. · Korean Journal of Odor Research and Engineering · 2013

The proficiency test programs for the air dilution olfactory method were carried out against 79 certification odor inspection agencies in Korea in 2011. Two synthetic proficiency testing materials (PTMs) which were simulating the complex odors at the site boundary line and the outlet of odor emission sources in the offensive odor control law were used. The boundary PTM was composed of 7ppm toluene and 7ppm m-xylene, and the outlet PTM was composed of 10ppm dimethylsulfide and 10ppm dimethyl disulfide. The results of the proficiency tests were evaluated by the Z-scores using mean or median as the assigned value and normal standard deviation, robust standard deviation, and coefficient of variation as the target standard deviation. Both the synthetic complex odors at the boundary and at the outlet were suitable as PTMS. The coefficient of variation of the test results was decreased with increasing the odor intensity of PTMs. Proficiency testing results for the complex odor was appropriate to assess using the odor index of the log scale than the odor dilution ratio. When Z-score for two PTMs was evaluate using the coefficient of variation standard deviations, and Robust standard deviation, 95% of the participants recorded the satisfactory performance. When the target standard deviation set up at 20% of the coefficient of variation of the test results, the rate of satisfactory performance for the boundary and outlet PTMs in the testing were 90% and 95%, respectively.