Вплив товщини рогівки на результати різних видів тонометрії

Ірина Миколаївна Безкоровайна, Денис Олександрович Наконечний · CyberLeninK (CyberLeninka) · 2014

134 eyes were examined, 67 students -volunteers of the medical higher educational establishment without any ophthalmological pathology. It was revealed that 60% of the examined eyes have the central corneal thickness which can essentially affect the results of intraocular pressure measurement. The measurement error between the data of Maklakovs tonometry and data of transpalpebral tonometry on the average is 0,5 mmHg in the first group, (1,0 mmHg) in the second group, (3,0 mmHg) in the third group, (3,8 mmHg) in the fourth group. It was found that an increase of frequency of error between the index obtained using different types of tonometry is directly proportional to the increase of corneal compensated ocular pressure due to increased thickness of the cornea.The most authentic at intraocular pressure are at Maklakovs tonometry, but this method is dependent on central thickness of a cornea. The last authentic of intraocular pressure are those obtained using the indicator IHD-02 PRA of transpalpebral tonometry, so this method does not depend on central thickness of a cornea.

Cited by 0 works.