Clinical Spirometry Using Computer Techniques
Stuart Warren Rosner, Alan R. Palmer, Steven A. Ward et al. · American Review of Respiratory Disease · 2015
The methodology for automatically measuring the component parameters of the forced expiratory spirogram using a programmed digital computer has been described previously (1). Since that time the automated system has been employed in two clinical trials (2, 3). These clinical experiences prompted several modifications, which are now in effect. A calibration system, which offers increased measurement accuracy has been designed for the bell-type spirometer and a coding system has been devised so that a subject's age, sex, height, and weight are introduced into the computer in advance of the spirographic signal. In addition, a diagnostic program has been developed which enables computer analysis to include an interpretation of the measurements as well as the computed values. This report describes the modified system and details the precision and accuracy of the measurements achieved with its use.
Cited by 1 works.
Authors
References
- Variability of the vital capacity of the normal human subject
- Methodology for Computer Measurement of Pulmonary Function Curves
- Measurement of the Mechanical Function of the Lung in Normal Subjects
- Observer Variation in Spirometry* *Presented in part at the 8th International Congress on Diseases of the Chest, Mexico City, October, 1964.
- Spirometry in a Medical Clinic