치과용 임플란트의 안전성과 유효성에 대한 드릴 디자인의 효과

윤지훈, 전계록, 윤미정 et al. · 대한치과의사협회지 · 2015

Purpose : The objective of this research was to investigate an optimization of drill design factors for implant stability and efficacy through comparative evaluation by the cutting time, heat generation and initial stability. Materials and Methods : Three design factors were considered for the purpose of drill design optimization; the number of flute(2 flute, 3 flute), helix angle(15°, 25°) and drill tip shape(straight, 2 phase). Design factors were selected through comparative evaluation by temperature change, cutting time and ISQ value. Results : Heat generation and cutting time are influenced by all design factors(p Conclusion : Drills with 2 flutes, 2 phase formed drill tip, and 25 degrees of helix angle exhibit a better performance than other design.