유량측정을 위한 초음파 전파경로 비교

김주완, 박춘광, 김진오 et al. · 한국소음진동공학회 학술대회논문집 · 2015

The paper deals with the accuracy comparison between two kinds of ultrasonic paths for flow rate measurement. In the Z-path, two transducers are installed on the opposite sides of each other on a pipe, and the ultrasonic waves generated at one transducer propagate to arrive at the other one only by refraction. In the V-path, two transducers are installed on one side of a pipe, and the ultrasonic waves reflected at the inner wall of the pipe are received. Transit times were confirmed to identify the propagation paths by comparing the theoretically calculated results and measured ones. The flow rate measurements with two kinds of ultrasonic paths appeared very similar. It would be possible to select either of the paths by considering the advantages and disadvantages.

Cited by 0 works.

Authors

References