A medical stimulator has been used for pain relief treatment. But, it cannot deliver stimuli to the target tissue without any affection of other tissues, and, moreover, the undesired stimuli to the patient's heart can cause dangerous heart failure. As a result, the current intensity of stimuli should be limited within proper values. There are available solutions that ensure safety and raise the efficiency of stimulation at a target tissues by designing electrodes and the stimulation waveform. In this paper, we measured and estimated the stimuli transmission through the electrolyte chamber that are simulating the patient's tissue at different conditions that have different values and waveforms and that are equipped with different electrodes. We tried to find the most optimal stimulation condition that can concentrate stimulation energy to the target tissue at various stimuli waveforms according to the design of electrodes. The basic design concept of electrode is that the circular reference electrode block off the leaking current from the stimulating electrode. And, the stimulation waveforms are monophasic and biphasic exponential truncated pulses.