Широкополосный измеритель биоимпеданса с адаптивным выбором сетки частот

V. S. Mosiychuk, Г. В. Тимошенко, Олег Борисович Шарпан · Вісник Національного технічного університету України Київський політехнічний інститут. Серія: Радіотехніка. Радіоапаратобудування · 2014

Introduction. For the diagnosis of functional state and structure of biological objects with weakly expressed irregularities it is important quickly and accurately to determine the ampli-tudeand phase-frequency characteristics. Therefore, the purpose of the article is a represen-tation of the results of the development of biological objects high-speed impedance meter with the ability to select adaptive grid measuring frequencies in the extended band. Structure of the impedance meter. Developed instrument is designed to measure the im-pedance of the object on four-electrode method. The device uses a frequency synthesizer am-plitude-phase detector integrally fabricated and active electrodes, by which the voltage con-trolled current source and the input buffer amplifiers with low input capacitance, are located in close proximity to the studied bioobject. This allowed to make measurements at frequencies up to 5 MHz. Instruments characteristics. To test the device characteristics the frequency characteris-tics of the test object (RC-chain) impedance were measured. It is composed of 5 precision resistors and capacitors. Parameters of the elements were measured preliminarily by labora-tory inductance, capacitance and resistance meter E7-12. The dependence of the measure-ment errors of the developed device in the frequency range from 1 kHz to 5 MHz is not more than 5% of the modulus of the impedance and not more than 2° of the phase.