다축 힘/모멘트 센서 교정기의 개발 및 그의 불확도 평가
김갑순, 윤정원 · Journal of the Korean Society for Precision Engineering · 2006
This paper describes the development of multi-axis force/moment sensor calibration system and its uncertainty evaluation. This sensor, which has maximum capacities of 2000N in forces and 400Nm in moments, can generate the continuous forces (±Fx, ±Fy and ±Fz) and moments (±Mx, ±My and ±Mz.). In order to be used multi-axis force/moment sensors in many kinds of industries, they are carried out the characteristic test or the calibration with the system that can generate the forces and the moments. The developed calibrating systems are the disadvantages of the low capacity, the generation of step force and moment with weights, the high coasts in manufacture and so on. In this paper, the multi-axis force/moment sensor calibration system that can generate the continuous forces and moments was developed. Its maximum capacities are 2000N in forces and 400Nm in moments. And the system was evaluated in the expanded relative uncertainty. They were ±0.0004 in forces ±Fx , ±Fy, ±Fz, and ±0.0004 in moments ±Mx, ±My, ±Mz.
Cited by 17 works.