벼 생체엽신 질소함량 측정을 위한 근적외선분광분석의 검량식 작성
권영립, 백미화, 최동칠 et al. · 2005
Near Infrared Spectroscopy (NIRS) has been used as a tool for the rapid, accurate and nondestructive assay of the fresh rice leaf in nitrogen content. NlRS used in this study was visible and near infrared spectroscopy type instrument, Foss model 6500. To obtain a useful calibration equation, standard regression between the data was analyzed by chemical analysis and by NIRS method. Accuracy of calibration equation for nitrogen content on fresh leaf of rice were 0.879, 0.858 and 0.819, respectively. Accuracy of calibration equation after outlier treatment increased as 0.017, 0.02 and 0.061 improved each with 0.896, 0.878 and 0.880, respectively. Calibration equation combined using merge function after accuracy of calibration equation more increased by 0.911. Difference analysis value between calibration equation and lab value by kjeldahl showed 0.001 %. With this as same result is the possibility of closing the deterioration of the sample in order to omit a construction and pulverization process it is judged with the fact that the nitrogen content measurement of the fresh rice leaf which the possibility of reducing an hour and an expense is by a near infrared spectroscopy technique will be possible.