전방십자인대 재건술에서 네비게이션 장치를 이용한 방법과 고식적인 방법에서 생체 내 등장성 비교
송은규, 선종근, 박상진 et al. · 2009
Purpose: To compare isometricity between the navigational and conventional technique during an anterior cruciate ligament (ACL) reconstruction before fixating the graft. Materials and Methods: Twenty-three patients with an ACL insufficiency were enrolled in this study. After finishing the tibial tunnel preparation using a navigational technique, a femoral tunnel site was marked using either the navigational (Orthopilot ® ) or conventional (Arthrex ® 5 ㎜ off-set guide) technique. After passing the suture material attached to the guide pin through each femoral site, isometric tests were then performed using a specially designed device during full range of motion of the knee. Results: An mean isometricity after the navigational and the conventional technique was 3.0 ㎜ and 4.6 ㎜ respectively. The isometricity after the navigational technique was significantly better than that after the conventional technique (p=0.000). The navigational isometric point was usually 0 to 5 ㎜ anterior or antero-lateral to that at the point of the conventional technique. The mean distance between the tunnels by two techniques was 1.35 ㎜. There was a strong positive correlation between the difference in isometicity and the positional relationship of the femoral tunnel (r²=0.778, p=0.000). Conclusion: An ACL reconstruction using the navigational technique provides better isometricity than that obtained using the conventional technique.
Cited by 0 works.