Медицинские магнитно-резонансные томографы КФТИ: технические характеристики, диагностические возможности, пекрспективы развития

К. М. Салихов, Я. В. Фаттахов, А. Р. Фахрутдинов et al. · CyberLeninK (CyberLeninka) · 2008

The medical magnetic resonance tomographs TMR-0.06-KFTI were developed and placed in service by Zavoisky Physical Technical Institute. This device can be used for diagnostics of brain, backbone and joints pathology. The device has following general technical and operating characteristics: magnetic field induction is 0.06 T; spatial resolution is no more than 1 mm; investigation layer's minimum thickness is 5 mm; maximum number of simultaneously obtained sections is 32; This tomograph successfully passed technical and medical acceptance tests that where carried out by leading specialists of Russian Ministry of Health. Members of inspection board decided that the magnetic resonance tomograph TMR-0.06-KFTI can be recommended for medical applications. It was confirmed that the technical and diagnostics characteristics of this device correspond to standard characteristics accepted in the world for low-field tomographs. At present in Zavoisky Physical-Technical Institute a new «whole body» model with magnetic field induction of 0.12 T, specialized joint tomographs with permanent and superconductor magnets are developed.