Validation of higher‐throughput high‐performance liquid chromatography/atmospheric pressure chemical ionization tandem mass spectrometry assays to conduct cytochrome P450s CYP2D6 and CYP3A4 enzyme inhibition studies in human liver microsomes
Inhou Chu, Leonard V. Favreau, Tony Soares et al. · Rapid Communications in Mass Spectrometry · 2000
In the early stage of drug discovery, thousands of new chemical entities (NCEs) may be screened before a single drug candidate can be identified for development. In order to accelerate the drug discovery process, we have developed higher-throughput enzyme assays to evaluate the inhibition of cytochrome P450 isoforms 2D6 (CYP2D6) and 3A4 (CYP3A4) in human liver microsomes. The assays are based on high-performance liquid chromatography/tandem mass spectrometry (LC/MS/MS) techniques. The analysis time for each sample was reduced from approximately 20 minutes for the conventional HPLC assay to 30 seconds for the LC/MS/MS assay. For both LC/MS/MS assays, the linearity (r(2) > 0.99), precision (%CV < 15%) and accuracy (% bias <15%) for both inter- and intraday validations were satisfactory. Since the implementation of the LC/MS/MS assays, our sample throughput has increased by over 40-fold.
Cited by 52 works. doi:10.1002/(sici)1097-0231(20000229)14:4<207::aid-rcm863>3.0.co;2-#
Authors
References
- Matrix Effect in Quantitative LC/MS/MS Analyses of Biological Fluids: A Method for Determination of Finasteride in Human Plasma at Picogram Per Milliliter Concentrations
- Mechanism-based enzyme inactivation : chemistry and enzymology
- Microtiter Plate Assays for Inhibition of Human, Drug-Metabolizing Cytochromes P450
- Steroid hormone hydroxylase specificities of eleven cDNA-expressed human cytochrome P450s
- The Simultaneous Determination of Mixtures of Drug Candidates by Liquid Chromatography/Atmospheric Pressure Chemical Ionization Mass Spectrometry as anIn Vivo Drug Screening Procedure
- Ketoconazole and sulphaphenazole as the respective selective inhibitors of P4503A and 2C9
- The need for adequate chromatographic separation in the quantitative determination of drugs in biological samples by high performance liquid chromatography with tandem mass spectrometry
- Application of a generic fast gradient liquid chromatography tandem mass spectrometry method for the analysis of cytochrome P450 probe substrates
- Simultaneous Determination of Dextromethorphan and Three Metabolites in Plasma and Urine Using High-Performance Liquid Chromatography with Application to Their Disposition in Man
- Measurement of Liver Microsomal Cytochrome P450 (CYP2D6) Activity Using [O-methyl-14C] Dextromethorphan