A Simple Statistical Parameter for Use in Evaluation and Validation of High Throughput Screening Assays
Ji-Hu Zhang, Thomas D.Y. Chung, Kevin R. Oldenburg · SLAS DISCOVERY · 1999
The ability to identify active compounds (³hits²) from large chemical libraries accurately and rapidly has been the ultimate goal in developing high-throughput screening (HTS) assays. The ability to identify hits from a particular HTS assay depends largely on the suitability or quality of the assay used in the screening. The criteria or parameters for evaluating the ³suitability² of an HTS assay for hit identification are not well defined and hence it still remains difficult to compare the quality of assays directly. In this report, a screening window coefficient, called ³Z-factor,² is defined. This coefficient is reflective of both the assay signal dynamic range and the data variation associated with the signal measurements, and therefore is suitable for assay quality assessment. The Z-factor is a dimensionless, simple statistical characteristic for each HTS assay. The Z-factor provides a useful tool for comparison and evaluation of the quality of assays, and can be utilized in assay optimization and validation.
Cited by 6,808 works. doi:10.1177/108705719900400206 Open-access PDF
Authors
References
- Applications of Combinatorial Technologies to Drug Discovery. 2. Combinatorial Organic Synthesis, Library Screening Strategies, and Future Directions
- Bacterial genomics
- The GDBTM Human Genome Database Anno 1997
- Assay Miniaturization for Ultra-High Throughput Screening of Combinatorial and Discrete Compound Libraries: A 9600-Well (0.2 Microliter) Assay System
- Toward a cDNA Map of the Human Genome
- Design of Signal Windows in High Throughput Screening Assays for Drug Discovery
- Combinatorial Organic Synthesis: Applications to Drug Discovery
- Comparison of Automation Equipment in High Throughput Screening