Повышение эффективности флуоресцентной диагностики новообразований кожи век и периорбитальной области эпителиального происхождения
Новиков Иван Александрович, Yaroslav O. Grusha, N. P. Kiryushchenkova · Вестник Российской академии медицинских наук · 2012
Developed in 2006 by our institute, the method of primary detection of eyelid and periorbital epithelial tumors based on non-induced fluorescence of endogenous protoporphyrin IX is used to estimate proliferative activity of tumor and its probable progression. This study was aimed to improve method efficacy by means of additional parameter which would quantitatively describe the level of tumor border dithering. The authors analyzed autofluorescent data of 97 patients and determined medico-biological factors which could cause controversal results (all cases were histologically controlled). The averaged reciprocal of tumor border probability was proposed as a solution and named the coefficient of tumor border dithering K djth. Repeated analysis of the autofluorescent images using both the proliferative factor and coefficient of interposition revealed improvement of method specificity from 69% to 85% while sensitivity of method remained 70%. In authors' opinion, obtained results can likely be extrapolated to neoplasm of other localization.