A multinational study to develop universal standardization of whole-body bone density and composition using GE Healthcare Lunar and Hologic DXA systems
J. Shepherd, Bo Fan, Ying Lü et al. · Journal of Bone and Mineral Research · 2012
Dual-energy x-ray absorptiometry (DXA) is used to assess bone mineral density (BMD) and body composition, but measurements vary among instruments from different manufacturers. We sought to develop cross-calibration equations for whole-body bone density and composition derived using GE Healthcare Lunar and Hologic DXA systems. This multinational study recruited 199 adult and pediatric participants from a site in the US (n = 40, ages 6 through 16 years) and one in China (n = 159, ages 5 through 81 years). The mean age of the participants was 44.2 years. Each participant was scanned on both GE Healthcare Lunar and Hologic Discovery or Delphi DXA systems on the same day (US) or within 1 week (China) and all scans were centrally analyzed by a single technologist using GE Healthcare Lunar Encore version 14.0 and Hologic Apex version 3.0. Paired t-tests were used to test the results differences between the systems. Multiple regression and Deming regressions were used to derive the cross-conversion equations between the GE Healthcare Lunar and Hologic whole-body scans. Bone and soft tissue measures were highly correlated between the GE Healthcare Lunar and Hologic and systems, with r ranging from 0.96 percent fat [PFAT] to 0.98 (BMC). Significant differences were found between the two systems, with average absolute differences for PFAT, BMC, and BMD of 1.4%, 176.8 g and 0.013 g/cm(2) , respectively. After cross-calibration, no significant differences remained between GE Healthcare Lunar measured results and the results converted from Hologic. The equations we derived reduce differences between BMD and body composition as determined by GE Healthcare Lunar and Hologic systems and will facilitate combining study results in clinical or epidemiological studies.
Cited by 218 works. doi:10.1002/jbmr.1654 Open-access PDF
Authors
References
- Measuring agreement in method comparison studies
- Improvements on Cross-Validation: The .632+ Bootstrap Method
- Universal standardization for dual X-ray absorptiometry: Patient and phantom cross-calibration results
- Dual Energy X-Ray Absorptiometry Body Composition Reference Values from NHANES
- Universal Standardization of Bone Density Measurements: A Method with Optimal Properties for Calibration Among Several Instruments
- Standardization of Bone Mineral Density at Femoral Neck, Trochanter and Ward's Triangle
- Precision and accuracy of measurements of whole-body bone mineral: comparisons between Hologic, Lunar and Norland dual-energy X-ray absorptiometers
- Universal standardization for dual X-ray absorptiometry: Patient and phantom cross-calibration results
- Anomalies in the Measurement of Changes in Total-Body Bone Mineral by Dual-Energy X-Ray Absorptiometry During Weight Change
- Anomalies in dual energy X-ray absorptiometry measurements of total-body bone mineral during weight change using Lunar, Hologic and Norland instruments.