Microprocessor-controlled photon-counting spectrofluorometer
Enrico Gratton, Mark Limkeman · Review of Scientific Instruments · 1983
The construction details and performance characteristics for a fluorescence spectrophotometer are described. The system utilizes proton counting for signal detection and an Apple II microcomputer for data acquisition and analysis. Data-acquisition techniques are given along with typical excitation, emission, and polarization spectra. Particular reference is made to application of the instrument to fluorescence studies of protein/polymer films.
Cited by 14 works. doi:10.1063/1.1137386
Authors
References
- The Fourier Transform and Its Applications
- Data smoothing using low-pass digital filters
- The measurement of luminescence waveforms by single-photon techniques
- Fluorescence polarization: measurements with a photon-counting photometer
- Construction and performance of a scanning, photon-counting spectrofluorometer
- Homogeneity and variability in the structure of azurin molecules studied by fluorescence decay and circular polarization
- The fine structure of luminescence spectra of azurin
- Data acquisition and control system based upon the Rockwell AIM-65 microcomputer
- Microcomputer fluorometer for corrected, derivative, and differential spectra and quantum yield determinations