High‐throughput liquid chromatography/mass spectrometry method for the quantitation of small molecules using accurate mass technologies in supporting discovery drug screening
Xiao Ding, Hesham Ghobarah, Xiaolin Zhang et al. · Rapid Communications in Mass Spectrometry · 2012
RATIONALE: Drug discovery samples are routinely analyzed using liquid chromatography/tandem mass spectrometry (LC/MS/MS) methods on triple quadrupole mass spectrometers employing multiple reaction monitoring (MRM). In order to improve analysis throughput, quantitation of small molecules on a quadrupole time-of-flight (QqTOF) instrument using TOF scan and high-resolution MRM (MRM-HR) modes was evaluated in this study. METHODS: Cassette dosed plasma and brain samples from nine compounds were extracted using a protein precipitation method. Separation was achieved by reversed-phase liquid chromatography. Mass spectrometric analysis was performed using TOF scan and high-resolution MRM approaches on a QqTOF mass spectrometer with turbo-ionspray ionization. Results were compared to those obtained on a triple quadrupole mass spectrometer. RESULTS: The dynamic range varied depending on compounds and instruments and was similar between the MRM on QqQ and full TOF scan mode on QqTOF. Linear or quadratic regression and 1/x(2) weighting were used. Resolution on the QqTOF instrument was around 32000 and mass accuracy was within 4.4 ppm. The MRM-HR method showed better sensitivity compared to the TOF scan method, and was comparable to the MRM on a QqQ mass spectrometer. Assay accuracy was within ±25%. CONCLUSIONS: A TOF scan method allowed the use of the generic method without compound-specific optimization and was an alternative choice for routine high-throughput quantitation of small molecules. The MRM-HR method on the QqTOF showed good sensitivity which was comparable to that obtained by the MRM method on the triple quadrupole mass spectrometer.
Cited by 26 works. doi:10.1002/rcm.6461
Authors
- Xiao Ding
- Hesham Ghobarah
- Allan Jaochico
- Xingrong Liu
- Gauri S. Deshmukh
- Bianca M. Liederer
- Cornelis E. C. A. Hop
- Brian Dean
References
- An introduction to quadrupole–time‐of‐flight mass spectrometry
- Quantification and Rapid Metabolite Identification in Drug Discovery Using API Time-of-Flight LC/MS
- In vivo pharmacokinetic screening in cassette dosing experiments: the use of on-line prospekt® liquid chromatography/atmospheric pressure chemical ionization tandem mass spectrometry technology in drug discovery
- Quantitative-qualitative data acquisition using a benchtop orbitrap mass spectrometer
- Atmospheric pressure ionization time-of-flight mass spectrometry coupled with fast liquid chromatography for quantitation and accurate mass measurement of five pharmaceutical drugs in human plasma
- Quantitation of small molecules using high‐resolution accurate mass spectrometers – a different approach for analysis of biological samples
- Rapid Determination of Oral Pharmacokinetics and Plasma Free Fraction Using Cocktail Approaches: Methods and Application
- Pharmacokinetic theory of cassette dosing in drug discovery screening.
- Targeted quantitative bioanalysis in plasma using liquid chromatography/high‐resolution accurate mass spectrometry: an evaluation of global selectivity as a function of mass resolving power and extraction window, with comparison of centroid and profile modes
- Full-scan high resolution accurate mass spectrometry (HRMS) in regulated bioanalysis: LC–HRMS for the quantitation of prednisone and prednisolone in human plasma