A fully automated optimization algorithm for determining the 3-D patient contour from photo-peak projection data in SPECT
Thomas J. Hebert, S.S. Gopal, Paul H. Murphy · IEEE Transactions on Medical Imaging · 1995
The class of noninverse problems versus inverse problems is discussed. A general optimization approach to solving certain constrained noninverse problems is presented. This approach is applied in the area of single photon emission computed tomography to estimate the patient outer contour directly from the photo-peak projection data. The resulting algorithm is fully automated, fast, and the determined 3-D patient contour satisfies smoothness constraints. The accuracy and reliability of the algorithm is evaluated through Monte Carlo simulations, patient studies, and an experimental validation study.
Cited by 14 works. doi:10.1109/42.370408
Authors
References
- Digital Image Processing
- A Method for Attenuation Correction in Radionuclide Computed Tomography
- Determination of Object Contour from Projections for Attenuation Correction in Cranial Positron Emission Tomography
- Compensation of tissue absorption in emission tomography
- Implementation of simultaneous attenuation and detector response correction in SPECT
- Nuclear Medicine and Biology 2009
- An attenuated projector-backprojector for iterative SPECT reconstruction
- Three-dimensional iterative reconstruction algorithms with attenuation and geometric point response correction
- Whole-body single-photon emission computed tomography using dual, large-field-of-view scintillation cameras
- Nonuniform attenuation correction using simultaneous transmission and emission converging tomography