This paper describes the design, fabrication, and test of a PDMS and PMMA -laminated microfluidic device for an immunosensing biochip. A PDMS top substrate molded by polymer casting and a PMMA bottom substrate fabricated by hot embossing are bonded with pressing force and hermetically sealed. Two inlet ports and an air vent are opened through the PDMS substrate, while Au electrodes for electrochemical detection are patterned onto the PMMA substrate. The analyte sample is loaded from the sample inlet port to the detection chamber by capillary force, without any intervening forces. For this and to control the time duration of sample fluid in each compartment of the device, including the inlet port, reaction barrier, reaction chamber, flow-delay neck, and the detection chamber, the fluid conduit has been designed using varying geometry. Especially, the fluid path has been designed so that the sample flow stops after filling the detection chamber to allow sufficient time for biochemical reaction and subsequent washing steps. The washing solution is supplied through the buffer solution inlet port by an external syringe pump. As a model system for a configurable microfluidic device, the functionalization of ferritin (antigen) to the sensing monolayer and their biospecific interaction with anti-ferritin antiserum has been investigated. An electrochemical detection method for immunosensing by biocatalyzed has been developed and applied for signal registration.