Multi-row linear cMUT array using cMUTs and multiplexing electronics
Robert Wodnicki, Charles Woychik, Albert Byun et al. · IEEE International Ultrasonics Symposium · 2009
A large area reconfigurable imaging array for research purposes is being developed with co-integrated cMUTs and control electronics. The goal is a 2.5cm 2D tileable module with >16,000 transducer sub-elements spaced at a pitch of 185um in X and Y dimensions. As a prototype demonstration of some of the goals of this effort, a multi-row linear array using cMUTs and external multiplexing electronics was designed and fabricated. In this paper the challenges of trenched cMUT attach to a laminate interposer as part of a tileable module will be discussed. The architecture of the tileable module build-up for manufacturability, reliability, acoustic planarity, and reduced spacing between tiles and cMUT chips will also be addressed. Finally, a first prototype will be shown and experimental acoustic results with the new cMUT-based probe will be presented.
Cited by 21 works. doi:10.1109/ultsym.2009.5442074
Authors
- Robert Wodnicki
- Charles Woychik
- Albert Byun
- Rayette Fisher
- Kai E. Thomenius
- Der-Song Lin
- Xuefeng Zhuang
- Ömer Oralkan
- Srikant Vaithilingam
- Butrus T. Khuri-Yakub
References
- Integration of 2D CMUT arrays with front-end electronics for volumetric ultrasound imaging
- Microfabricated ultrasonic transducers monolithically integrated with high voltage electronics
- Wafer-bonded 2-D CMUT arrays incorporating through-wafer trench-isolated interconnects with a supporting frame
- Reconfigurable arrays for portable ultrasound
- 5F-2 Packaging and Design of Reconfigurable Arrays for Volumetric Imaging