Microwave Ablation with Multiple Simultaneously Powered Small-gauge Triaxial Antennas: Results from an in Vivo Swine Liver Model
Christopher L. Brace, Paul F. Laeseke, Lisa A. Sampson et al. · Radiology · 2007
PURPOSE: To prospectively investigate the ability of a single generator to power multiple small-diameter antennas and create large zones of ablation in an in vivo swine liver model. MATERIALS AND METHODS: Thirteen female domestic swine (mean weight, 70 kg) were used for the study as approved by the animal care and use committee. A single generator was used to simultaneously power three triaxial antennas at 55 W per antenna for 10 minutes in three groups: a control group where antennas were spaced to eliminate ablation zone overlap (n=6; 18 individual zones of ablation) and experimental groups where antennas were spaced 2.5 cm (n=7) or 3.0 cm (n=5) apart. Animals were euthanized after ablation, and ablation zones were sectioned and measured. A mixed linear model was used to test for differences in size and circularity among groups. RESULTS: Mean (+/-standard deviation) cross-sectional areas of multiple-antenna zones of ablation at 2.5- and 3.0-cm spacing (26.6 cm(2) +/- 9.7 and 32.2 cm(2) +/- 8.1, respectively) were significantly larger than individual ablation zones created with single antennas (6.76 cm(2) +/- 2.8, P<.001) and were 31% (2.5-cm spacing group: multiple antenna mean area, 26.6 cm(2); 3 x single antenna mean area, 20.28 cm(2)) to 59% (3.0-cm spacing group: multiple antenna mean area, 32.2 cm(2); 3 x single antenna mean area, 20.28 cm(2)) larger than 3 times the mean area of the single-antenna zones. Zones of ablation were found to be very circular, and vessels as large as 1.1 cm were completely coagulated with multiple antennas. CONCLUSION: A single generator may effectively deliver microwave power to multiple antennas. Large volumes of tissue may be ablated and large vessels coagulated with multiple-antenna ablation in the same time as single-antenna ablation.
Cited by 118 works. doi:10.1148/radiol.2441052054
Authors
- Christopher L. Brace
- Paul F. Laeseke
- Lisa A. Sampson
- Tina M. Frey
- Daniel Warren van der Weide
- Fred T. Lee
References
- Tissue ablation with radiofrequency using multiprobe arrays.
- Hepatic Microwave Ablation With Multiple Antennae Results in Synergistically Larger Zones of Coagulation Necrosis
- Effect of Vascular Occlusion on Radiofrequency Ablation of the Liver
- Heating characteristics of array applicator composed of two coaxial-slot antennas for microwave coagulation therapy
- Microwave ablation with a triaxial antenna: results in ex vivo bovine liver
- Microwave Ablation with a Single Small-Gauge Triaxial Antenna: In Vivo Porcine Liver Model
- Multiple Probe Radiofrequency Ablation: Pilot Study in an Animal Model
- Radiofrequency Ablation of Hepatic Tumors: Variability of Lesion Size Using a Single Ablation Device
- Multiple-Electrode Radiofrequency Ablation Creates Confluent Areas of Necrosis: In Vivo Porcine Liver Results
- Large-Volume Radiofrequency Ablation of ex Vivo Bovine Liver with Multiple Cooled Cluster Electrodes