Discriminating Between Normal and Glaucomatous Eyes Using the Heidelberg Retina Tomograph, GDx Nerve Fiber Analyzer, and Optical Coherence Tomograph
Linda M. Zangwill · Archives of Ophthalmology · 2001
OBJECTIVE: To compare the ability of 3 instruments, the Heidelberg Retina Tomograph (HRT), the GDx Nerve Fiber Analyzer (GDx), and the Optical Coherence Tomograph (OCT), to discriminate between healthy eyes and eyes with early to moderate glaucomatous visual field loss. SUBJECTS AND METHODS: Forty-one patients with early to moderate glaucomatous visual field loss and 50 healthy subjects were included in the study. The HRT, GDx, and OCT imaging and visual field testing were completed on 1 eye from each subject within a 6-month interval. Statistical differences in sensitivity at fixed specificities of 85%, 90%, and 95% were evaluated. In addition, areas under the receiver operating characteristic (ROC) curve were compared. RESULTS: No significant differences were found between the area under the ROC curve and the best parameter from each instrument: OCT thickness at the 5-o'clock inferior temporal position (mean +/- SE, 0.87 +/- 0.04), HRT mean height contour in the nasal inferior region (mean +/- SE, 0.86 +/- 0.04), and GDx linear discriminant function (mean +/- SE, 0.84 +/- 0.04). Twelve HRT, 2 GDx, and 9 OCT parameters had an area under the ROC curve of at least 0.81. At a fixed specificity of 90%, significant differences were found between the sensitivity of OCT thickness at the 5-o'clock inferior temporal position (71%) and parameters with sensitivities less than 52%. Qualitative assessment of stereophotographs resulted in a sensitivity of 80%. CONCLUSION: Although the area under the ROC curves was similar among the best parameters from each instrument, qualitative assessment of stereophotographs and measurements from the OCT and HRT generally had higher sensitivities than measurements from the GDx.
Cited by 344 works. doi:10.1001/archopht.119.7.985
Authors
References
- Comparing the Areas under Two or More Correlated Receiver Operating Characteristic Curves: A Nonparametric Approach
- Optic Nerve Damage in Human Glaucoma
- Quantification of Nerve Fiber Layer Thickness in Normal and Glaucomatous Eyes Using Optical Coherence Tomography
- The Nerve Fiber Layer in the Diagnosis of Glaucoma
- Clinically Detectable Nerve Fiber Atrophy Precedes the Onset of Glaucomatous Field Loss
- Optic Nerve Damage in Human Glaucoma
- Reproducibility of Nerve Fiber Layer Thickness Measurements Using Optical Coherence Tomography
- The Mode of Progressive Disc Cupping in Ocular Hypertension and Glaucoma
- Optic Disc Parameters and Onset of Glaucomatous Field Loss
- Scanning Laser Polarimetry to Measure the Nerve Fiber Layer of Normal and Glaucomatous Eyes