Аппаратура для флюоресцентной диагностики воспалительных заболеваний полости рта
Anatoly N. Berezin, Maria L. Sinyaeva, Sergey A. Shevchik et al. · Российский биотерапевтический журнал · 2004
Devices and method of acquiring parodontal autofluorescent images is presented in this work. The first device is destined to register autofluorescence with low intensity that appears from pathological microflora. Device high sensibility is achieved with presence in the excitation spectra both violet and red component. This allows to excite big intensity fluorescence on the surface with radiation on 380530 nm and in the depth of tissue with radiation on 640?50 nm. Autofluorescence registration is made on 690+ nm. IOK-01-Biospec device is destined both for registration of microflora either teeth enamel autofluorescence. It is equipped with light emitting diodes and registration filter. With the help of these devices 14 patients with parodontal tissue inflammatory disease and demineralization enamel nidus were examined. Autofluorescence visual data fits with spectroscopic investigation data carried out with spectrometer LESA-01-Biospec. Research results have demonstrated the efficacy of developed devices and FD method for demineralization enamel nidus and inflammatory disease localization.