집속형 초음파 조사 시 천이형 캐비테이션 발생이 혈액에 미치는 영향에 관한 연구

민정화, 정현두, 김진수 et al. · 대한전자공학회 학술대회 · 2012

This paper is a basic study to evaluate the vascular effect of focused ultrasound. To reproduce an environment similar to the human body, blood flow mimicking model of the dynamic state was manufactured by using TMM phantom and latex tube. The ultrasound pulses were produced by a 500 kHz concave type focused transducer. The experimental groups were divided into two groups UCA injection and UCA non-injection and the exposure was conducted in the range of low-intensity ultrasound at different intensity and a pixed duration time for 3-mins. Once the exposure is complete, blood sample is diluted at 1:200 ratio by using Gower solution. Red blood cells were counted by using a hemacytometer and viscosity was measured. The results of experiment show that there is no significant difference between the two groups and blood cells and viscosity were decreased with the increasing intensity.

Cited by 0 works.