시-주파수 분석을 이용한 심실세동시 심전도 분석을 통한 제세동 예측에 관한 연구
成洪模, 申在佑, 李賢淑 et al. · 전기학회논문지 D · 2006
The purpose of this study is to predict the defibrillation success of a ventricular Fibrillation ECG signal using time-frequency analysis. During CPR, coronary perfusion pressure and electrocardiogram were measured. Parameters extracted from time-frequency domain were served as predictor of resuscitation success. Time frequency distribution(TFD) of ECG signals was estimated from the smoothed pseudo Wigner-Ville distribution(SPWVD). Median frequency, peak frequency, 1/f slope, frequency band ratios(2~4㎐, 4~6㎐, 6~8㎐, 8~10㎐, 10~12㎐, 12~15㎐) were extracted from each TFD as function of time. Paired t-test was used to determine the differences in ROSC and non-ROSC groups. In the statistical results, we selected four significant parameters - median frequency, 1/f slope, 2~4㎐ band ratio, 8~10㎐ band ratio. We made an attempt to predict defibrillation success by combining features extracted from time frequency distribution. Independent t-test was used to determine the differences ROSC and non-ROSC groups. Consequently, we selected four significant parameters - median frequency, 1/f slope, 2~4㎐ band ratio, 8~10㎐ band ratio. The relationship between coronary perfusion pressure and ECG parameters was analyzed with linear regression analysis. R-square value was 55%. 1/f slope and 8~10㎐ band ratio had the significant relationship with coronary perfusion pressure.
Cited by 0 works.