이중 헥사포드 외고정 기구를 이용한 선천성 왜소증 환자의 경골 교정 시뮬레이션
주상민, 김윤혁 · 한국정밀공학회 학술발표대회 논문집 · 2004
In this paper, we proposed a new method of pre-operative planning for tibial deformity correction using double hexapod external fixator in pseudoacondroplasia. The 3-D computer graphic model of deformed tibia was reconstructed from 3 ㎜ sliced CT data, and CAD model of double hexapod external fixator was developed. The fixator was composed of 170 ㎜ diameter of three rings and 90 ㎜ of twelve struts. The bone deformities and the osteotomy lines for double osteotomy were measured using X-rays, and the necessary joint values to correct the given deformities were obtained by inverse kinematics analysis. The computer graphic simulation was performed to visualize the deformity correction process and evaluate the analysis result. By examining the pre-op and post-op X-rays, the simulation result was in good agreement with the clinical outcomes.
Cited by 0 works.