시뮬레이션 해석에 따른 초음파 변환자 설계의 정확도 평가

박찬육, 성진호, 정종섭 · 대한전자공학회 학술대회 · 2014

In this paper, we evaluated the accuracy of the ultrasound transducer design based on KLM/FEM analysis and fabrication of prototype device. A 8 MHz single element transducer and 6.5 MHz linear array transducer were designed with PiezoCAD (KLM) and PZFlex (FEM), subsequently, prototype transducers were fabricated. The performance of simulation and experimental results including the center frequency and ?6 dB bandwidth obtained by pulse-echo measurement was compared to each other. The center frequency and ?6 dB bandwidth of simulated and fabricated devices were similar. There was some difference between simulation and experiment due to fabrication and measurement environment errors, however, that can be overcome by fabricating and measuring more carefully.

Cited by 0 works.