Electromagnetic waves could have destructive effects on people’s health. In modern war strategies, electromagnetic methods are used to disable the man power. Radiation of electromagnetic fields could change some physiological parameters of the body, such as blood glucose level. So it needs frequent blood glucose measurement by sampling. In this article, we have proposed a new non-invasive method for blood glucose measurement. This electromagnetic method could be used for measurement on the tissue of a living creature. In this method, a bobbin has been used as a sensor, which the solution that contains the glucose (i.e. blood) is poured in its core. Any change in glucose concentration causes the dielectric solution parameter variation and hence the impedance of the bobbin changes. The ratio of the secondary bobbin signal to the reference signal of the primary bobbin, T/R ratio, has been measured in RF frequencies and used as an index of circuit impedance variation. Normally we expect that T/R would only change as a function of glucose, but there are other interference parameters such as the solution volume which affect our measurements. In this paper, we have proposed the measurement setup and the method of changing glucose concentration. We show that the volume variation has higher effects on our measurement output compared to that of glucose. We have proposed a method for compensation of the volume changes. Finally, we have shown that after this compensation, there will be a good correlation between the glucose concentration and the T/R ratio.