국제표준에 대비한 국내 비자동저울의 품질평가에 관한 연구

남궁재관, 공재향 · 한국공작기계학회 논문집 · 2004

As the result of reduction of the barrier of national economy on matters of quality assurance of weighing instruments in recently, it is considering in domestic as well as international matters. Therefore, this study is to analysis and compare with international and national measuring standards on the electric self-indicating scale, to improve the quality of electronic machine by providing a reformation plan on currently problems of domestic manufactured, and to identify to introduce in domestic criteria from international standards. I had compared KS C 1313 to OIML R 76-1 as the performance assessment items, such as metrology properties test, static temperature test, temperature effect testing under the un-load, electronic wave disturbance test, high-temperature and high-humidity stability test and durability test. Therefore, only one company is passed all items of the test according to the international (reference) standards, but it is possible to improve the quality in general if it should be use the load cell and electric components which is stability of temperature change in order to supplementation to the static temperature test and temperature effect testing under the un-load. It is also possible to apply in the domestic with the OIML 76-1 after correcting the design.

Cited by 1 works.

Authors