도플러 레이더를 이용한 비접촉식 무호흡 검출
조성필, 송미혜, 이정훈 et al. · 대한전자공학회 학술대회 · 2011
In this paper, to identify the feasibility of respiration monitoring using the Doppler radar, the apnea detection was demonstrated. For this study, the apnea mimicking protocol was devised and data were collected from 13 volunteers. The dual thresholds and instantaneous duration based algorithm was developed to detect the apneic event. In order to assess the performance, sensitivity and positive predictive value (PPV) were employed. The sensitivity and PPV were 97.04% and 99.49%, respectively.
Cited by 0 works.