좌심실보조장치의 출구 캐뉼라의 삽관 위치에 따른 대동맥궁 내부의 혈류 수치해석

지인혁, 임기무 · 대한기계학회 춘추학술대회 · 2014

Outflow cannulation site of left ventricular assist device (LVAD) chosen by considering anatomical structure of thoracic cavity and vascular system. Though outflow cannulation site influences blood perfusion at each branch, there is no standard rule or quantitative data. In this study, we computed the amount of blood perfusion at each arterial branch numerically according to outflow cannulation sites (ascending aorta, aortic arch, descending aorta). We generated computational meshes to the three-dimensionally reconstructed arterial system. Clinically measured arterial pressure were used for inlet boundary condition, porous media were applied to mimic blood flow resistance. Blood perfusion through left common carotid artery was 2.5 times higher than other cases, and that through right common carotid artery was 1.1 times higher than other branches. Although this is simulation study, will be useful reference data for the clinical study of LVAD which considers blood perfusion efficiency.